< Back
other-ringversuch

2018 - nanoanalytics - Quantitative X-ray microanalysis by SEM/EDX

Organizer
NanoAnalytics, Münster
Material
Laboratory comparison EDX (LV18) - EQAS test by nanoanalytics, Münster for quantitative X-ray microanalysis using SEM/EDX acc. to ISO 22309 on Y-Ni-Si alloy
Literature
Report on Laboratory Comparison EDX (LV18)
Download

Major elements [wt %]

CRB
RV
1sRV
Z-Score
Ni
40.300
43.900
3.200
-1.000
Y
36.400
35.000
2.600
0.460
Si
23.300
21.100
2.300
0.920

Legende

CRB
Interlaboratory test
1s-RV
Z-score
*
CRB Results
Interlaboratory Test Results
Standard Deviation Interlaboratory Test
Difference between the measured value and the average value of the interlaboratory test
Value not certified