< Back
other-ringversuch

2019 - nanoanalytics - Quantitative X-ray microanalysis by SEM/EDX

Organizer
NanoAnalytics, Münster
Material
Laboratory comparison EDX (LV19) - EQAS test by nanoanalytics, Münster for quantitative X-ray microanalysis using SEM/EDX acc. to ISO 22309 on a chromium-nickel steel
Literature
Report on Laboratory Comparison EDX (LV19)
Download

Major elements [wt %]

CRB
RV
1sRV
Z-Score
Fe
65.200
65.700
0.770
-0.600
Cr
18.300
17.600
0.620
1.150
Ni
11.200
12.200
0.410
-2.340
Mo
2.850
2.610
0.330
0.730
Mn
1.370
1.260
0.370
0.300
Si
0.430
0.410
0.100
0.200
Cu
0.180
0.270
0.090
-1.000
V
0.090
0.130
0.100
-0.400
Co
0.350
0.600
0.300
-0.830

Legende

CRB
Interlaboratory test
1s-RV
Z-score
*
CRB Results
Interlaboratory Test Results
Standard Deviation Interlaboratory Test
Difference between the measured value and the average value of the interlaboratory test
Value not certified