< Back
other-ringversuch
2022 - nanoanalytics - Quantitative X-ray microanalysis by SEM/EDX - 2022
Organizer
NanoAnalytics, Münster
Material
Laboratory comparison EDX (LV22) - EQAS test by nanoanalytics, Münster for quantitative X-ray microanalysis using SEM/EDX acc. to ISO 22309 on a chromium-nickel steel
Literature
Report on Laboratory Comparison EDX (LV22
Download
Major elements [wt %]
CRB
RV
1sRV
Z-Score
Si
0.680
0.700
0.220
0.090
V
0.060
0.050
0.020
-0.500
Cr
17.330
18.000
0.500
1.340
Mn
1.050
1.100
0.200
0.250
Fe
72.340
72.400
0.520
0.120
Ni
7.080
7.100
0.240
0.080
Cu
0.220
0.060
0.300
-0.530
Mo
0.360
0.300
0.120
-0.500
Legende
CRB
Interlaboratory test
1s-RV
Z-score
*
CRB Results
Interlaboratory Test Results
Standard Deviation Interlaboratory Test
Difference between the measured value and the average value of the interlaboratory test
Value not certified